Instrument Specifications: | |
Temperature Range | 10 °C to 90 °C |
Humidity Range | 5% to 95% Rh |
Can Dimensions | 3.00" ID, 2.6" tall |
Can Windows | Aluminum, 2.00" tall, 0.020" thick |
The SANS humididty chamber is a general purpose humidity generator that can control relative humidity from 5% to 95% Rh in the temperature range of 10 °C to 90 °C. The humidity is controlled by heating and cooling water wells within the chamber to generate or remove humidity from the cell.
The sample temperature is controlled by circulating ethylene glycol and the wells are heated and cooled by peltier heaters controlled by a bipolar power supply. To promote condensation on the sample, the outer can is heated to above the sample temperature using band heaters.
| SANS Home | | SANS Instruments | | Sample Environments | | SANS Publications | | Instrument Schedules | | Low-Q Seminar Series | | SANS Staff | | Data Reduction and Analysis | | Tutorials | | Info For Users | | SANS Calculations |
Last modified 06-November-2014 by website owner: NCNR (attn: Jeff Krzywon)